Effect of mechanical fatigue on commercial bioprosthetic TAVR valve mechanical and microstructural properties
Valvular structural deterioration is of particular concern for transcatheter aortic valve replacements due to their suspected shorter longevity and increasing use in younger patient populations. In this work we investigated the mechanical and microstructural changes in commercial TAVR valves composed of both glutaraldehyde fixed bovine and porcine pericardium (GLBP and GLPP) following accelerated wear testing (AWT) as outlined in ISO 5840 standards. This provided greater physiological relevance...